Horiba Bromide Ion Electrode – 8005-10C
This electrode determines the bromide ion concentration in aqueous solutions. The measurement range is 0.8 to 80,000 mg/L Br-.
Horiba Bromide Ion Electrode – 8005-10C
The Horiba Bromide Ion Electrode is great for determining the bromide ion concentration in aqueous solutions. It requires the 2565A-10T reference electrode which can be found on our site. It works with the majority of Horiba water testing meters and provides highly accurate results.
Horiba sensors and electrodes are among the highest quality in the world. They are made with top quality components and are designed to provide accurate, dependable results.
Specifications:
Measuring range |
0.8 to 80,000 mg/L Br- |
pH range |
800 mg/L (10-2 mol/L) Br- |
Applicable temperature range |
0 to 50℃ |
Response time (90%) |
Within 5 seconds |
Applicable reference electrode |
2565A |
Selection coefficient |
S2O32-、l-、S2-、CN- = Not acceptable MnO4- = 1 Cl-、PO42- = 100 F-、NO3-、SO42- = 1,000 |
• The response time is the time which is required to reach 90% response when the ion concentration is gradually changed from 10-4 mol/L to 10-2 mol/L with the solution stirred.
Exception:
*1: 90% response when ion concentration is changed to 10-6 mol/L ∼ 10-2 mol/L
*2: 95% response when ion concentration is changed to 10-3 mol/L ∼ 10-1 mol/L
*3: 95% response when ion concentration is changed to 10-4 mol/L ∼ 10-2 mol/L
*4: 95% response when ion concentration is changed to 10-4 mol/L ∼ 10-1 mol/L
• The selection coefficient is a ratio of the limit concentration of coexisting ions (mol/L) to the ion concentration to be measured (mol/L); The value of 1000 means that the coexisting ions can be permitted up to 1000 times the ion measured and ''not acceptable'' means that chemical change occurs in the solid response membrane.
Note: This electrode does not work with the Horiba D-53 meter.
Unit Ship Weight: 1 lb