Horiba Thiocyanate Ion Electrode – 8009-10C
This electrode determines the thiocyanate ion concentration in aqueous solutions. It has a measurement range of 0.6 to 5,800 mg/L SCN-.
Horiba Thiocyanate Ion Electrode – 8009-10C
The Horiba Thiocyanate Ion Electrode determines the thiocyanate ion concentration in aqueous solutions. It requires a 2565A-10T reference electrode to function properly and can be used with the majority of Horiba water testing meters.
Horiba sensors and electrodes are among the highest quality in the world. They are made with top quality components and are designed to provide accurate, dependable results.
Specifications:
Measuring range |
0.6 to 5,800 mg/L SCN- |
pH range |
5.8 mg/L (10-4 mol/L) SCN- |
Applicable temperature range |
0 to 50℃ |
Response time (90%) |
Within 30 seconds |
Applicable reference electrode |
2565A |
Selection coefficient |
CN-、 l-、 S2-、 S2O32- = Not acceptable Br- = 1 Cl- = 100 |
• The response time is the time which is required to reach 90% response when the ion concentration is gradually changed from 10-4 mol/L to 10-2 mol/L with the solution stirred.
Exception:
*1: 90% response when ion concentration is changed to 10-6 mol/L ∼ 10-2 mol/L
*2: 95% response when ion concentration is changed to 10-3 mol/L ∼ 10-1 mol/L
*3: 95% response when ion concentration is changed to 10-4 mol/L ∼ 10-2 mol/L
*4: 95% response when ion concentration is changed to 10-4 mol/L ∼ 10-1 mol/L
• The selection coefficient is a ratio of the limit concentration of coexisting ions (mol/L) to the ion concentration to be measured (mol/L); The value of 1000 means that the coexisting ions can be permitted up to 1000 times the ion measured and ''not acceptable'' means that chemical change occurs in the solid response membrane.
Note: This electrode does not work with the Horiba D-53 meter.
Unit Ship Weight: 1 lb